Question

Question Posted 08/21/15:
Dear sir/madam, I wished to know the utility of dynamic (six 5 min frames) acquisition over static for FDG-PET
Response posted 08/21/15 by Robert Koeppe:
This in entirely to help prevent the loss of a PET scan due to subject motion. In the processing steps, the individual 5 min frames are realigned to one another. In addition, if one or more frames has severe motion or severe attenuation correction artifacts due to mis-registration with the CT or transmission scan, these "bad" frames can be failed, while the rest of the frames can be passed and used for analysis.
Go back to list of topics >