  • Select the area you would like to search.
  • ACTIVE INVESTIGATIONS Search for current projects using the investigator's name, institution, or keywords.
  • EXPERTS KNOWLEDGE BASE Enter keywords to search a list of questions and answers received and processed by the ADNI team.
  • ADNI PDFS Search any ADNI publication pdf by author, keyword, or PMID. Use an asterisk only to view all pdfs.

Question

Question Posted 10/24/16:
I noticed, in the data collections, that different specifications are written regarding the scans : IR-SPGR, wacceleration calibration scan, 3-plane localizer, MPRAGE, MPRAGE GRAPPA2, MPRAGE SENSE 2, ... If I want the MPRAGE images, should I only download the images where the description MPRAGE is written ? What is the difference between MPRAGE and MPRAGE SENSE 2 or MPRAGE GRAPPA 2? Is there a file describing and comparing these specifications ?
Response posted 10/24/16 by ADNI MRI Core:
Thank you for your question to the ADNI MRI Core.

The difference between those series is that SENSE (Philips) and GRAPPA (Siemens) are factor of 2 accelerated scans and just MPRAGE is non accelerated.

Yes, If you wished only for non accelerated scans you should download series with only MPRAGE in the series description.

Download questions can be directed to LONI at: dba@loni.usc.edu

Thank you

ADNI MRI

Go back to list of topics >